Application of Ultrasonic C-scan in the Brazing Quality ofC/ SiC Composite and Titanium Alloy
DOI:
Author:
Affiliation:

Clc Number:

TB553

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    The sound impedance of C/ SiC composite is measured by the method of ultrasonic immersing pulse
    echo,and which is between 4. 0×10 6 kg/ m 2 s and 4. 7×10 6 kg/ m 2 s. The relationship of reflection is about 1 when there
    is crack in brazing interface and it is between 0. 71 and 0. 74 when there is no crack. For the differential value of the
    amplitude of reflection wave is at least 6 dB between two status, the amplitude of more than second reflection wave
    will be monitored when C-scan. The result of C-scan is certified by metallographic analysis, and the uniformity of both
    is good.